ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ ASPHERIC IOL WITH THE ULTRASERT PRE-LOADED DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number AU00T0
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare professional reported during an intraocular lens implant procedure, the lens stuck in cartridge.There was patient contact but no harm was reported.The lens was replaced with a back up lens in the initial procedure.Additional information has been requested.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been one other complaint reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was not returned for analysis.The complainant indicates the use of non company viscoelastic, which is not qualified for use with associated device.While we are unable to determine the origin of the reported complaint, our observations reasonably suggest that it is not manufacturing related.Based on the information provided by the customer, the most likely root cause is failure to follow ifu, as the customer states the use of non-qualified viscoelastic.The use of an unqualified ovd may cause damage to the lens and potential complications during the implantation process.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC IOL WITH THE ULTRASERT PRE-LOADED DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13503990
Report Number9612169-2022-00053
Device Sequence Number3953125
Product Code HQL
UDI-Device Identifier00380652394864
UDI-Public00380652394864
Combination Product (Y/N)N
Initial Reporter CountryNO
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/25/2024
Device Model NumberAU00T0
Device Lot Number25168836
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/17/2022
Supplement Date Received by Manufacturer04/06/2022
Initial Report FDA Received Date02/10/2022
Supplement Report FDA Received Date05/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MICROVISC PLUS
Patient Age74 YR
Patient SexFemale
-
-