Model Number 1884012EM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis results were not available as of the date of this report.A follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during fess surgery, connection from the blade is broken near the hub.There was no intervention performed / planned.No broken pieces of the reported product remain inside the patient's body.There was no delay in procedure.The procedure was completed with backup product(s).
 
Additional Manufacturer Narrative
H3 : analysis of part number 1884012em of lot number 0222708678 found that there was a residue consistent with biological contaminants on the device.Portions of the tracker wire were cut.Visually, the inner shaft broke 0.57 inches from the distal end of the inner hub which would have resulted in the reported event.There were striations around the outside diameter of the break point.Under magnification, there were indentations on the rotating hub.Functional testing could not be performed due to the broken state of the device.There was no allegation of a defect prior to use.A review of the global complaint data showed no other complaints about this lot number.In the returned condition, there was an out of specification condition that was related to the complaint due to physical damage.H6 : previous fdm b21, fdr c21, fdc d16 and img g04041 codes no longer apply.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
On follow-up it was reported that the blade broken while attempting to remove the blade from the handpiece to clear the blockage.Fragment detached only when the shaver blade has been removed from the handpiece to clean the blade and broken parts contained in the handpiece or in the outer blade that prevented the fragments from contacting the patient.There has been no patient impact, but the operation took longer since a new shaver blade had to be obtained and connected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS® SYSTEM NAVIGATED BLADE
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer (Section G)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
Manufacturer Contact
glen belmer
6743 southpoint drive north
jacksonville, FL 32216
6122713209
MDR Report Key13504806
Report Number1045254-2022-00056
Device Sequence Number12100832
Product Code EQJ
UDI-Device Identifier00613994617415
UDI-Public00613994617415
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/25/2023
Device Model Number1884012EM
Device Catalogue Number1884012EM
Device Lot Number0222708678
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2022
Supplement Date Received by Manufacturer02/14/2022
Initial Report FDA Received Date02/10/2022
Supplement Report FDA Received Date03/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-