MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA GMK-PRIMARY TIBIAL INSERT PS FIXED SIZE 3/10MM; KNEE TIBIAL INSERT PE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA GMK-PRIMARY TIBIAL INSERT PS FIXED SIZE 3/10MM; KNEE TIBIAL INSERT PE Back to Search Results
Model Number 02.07.F11066
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Batch review performed on 07 february 2022.Lot 175534: (b)(4) items manufactured and released on 21-dec-2017.Expiration date: 2022-11-21.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
 
Event or Problem Description
The extension stem screw broke during surgery.The surgeon had to remove the cemented tibial tray on the patient to change it to another cemented tibial tray, the liner was also changed.The locking screwdriver used was the torque limiter wrench of 6 nm.
 
Additional Manufacturer Narrative
Batch review performed on 07 february 2022.Lot 175534: (b)(4) items manufactured and released on 21-dec-2017.Expiration date: 2022-11-21.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
 
Event or Problem Description
The extension stem screw broke during surgery.The surgeon had to remove the cemented tibial tray on the patient to change it to another cemented tibial tray, the liner was also changed.The locking screwdriver used was the torque limiter wrench of 6 nm.
 
Additional Manufacturer Narrative
Note: initially it was reported that the extension stem screw broke.Finally, it was discovered that liner screw broke during surgery.Batch review performed on 12 april 2022.Lot 2106049: (b)(4) items manufactured and released on 20-aug-2021.Expiration date: 2026-0721.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Visual inspection performed by r&d project manager.The fixation screw of the gmk primary ps tibial insert broke in the attempt of screwing into the tibial tray.As reported in the event description the surgeon has wrongly used the 6nm torque limiter screwdriver (ref.02.07.10.4563).The applied torqued has exceeded the maximum torque that the screw can withstand, causing the breakage.The gmk primary surgical technique prescribes to use of the 3.5nm torque limiter screwdriver (02.07.10.4577) to guarantee that the optimal locking of the screw is achieved.Based on the available information, it's possible to assert that the event has not been caused by a faulty device.
 
Event or Problem Description
The liner screw broke during surgery.The surgeon had to remove the cemented tibial tray on the patient to change it to another cemented tibial tray, the liner was also changed.The locking screwdriver used was the torque limiter wrench of 6 nm (ref: 02.07.10.4563).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-PRIMARY TIBIAL INSERT PS FIXED SIZE 3/10MM
Common Device Name
KNEE TIBIAL INSERT PE
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key13505001
Report Number3005180920-2022-00077
Device Sequence Number14790730
Product Code JWH
UDI-Device Identifier07630030817069
UDI-Public07630030817069
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K090988
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/21/2026
Device Model Number02.07.F11066
Device Catalogue Number02.07.0310PSF
Device Lot Number2106049
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer01/12/2022
Initial Report FDA Received Date02/10/2022
Supplement Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age70 YR
Patient SexFemale
Patient Weight60 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-