• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE 100 ACET CUP 54MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE 100 ACET CUP 54MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-01-054
Medical Device Problem Codes Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Pain (1994); Ambulation or Postural Difficulties (2544); Joint Laxity (4526)
Date of Event 01/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Clinical adverse event received for dislocation.Event is serious and is considered moderate.Event is possibly related to device and is possibly related to procedure.Date of implantation: (b)(6) 2016, date of event: (b)(6) 2022, (right hip).Treatment: closed reduction/manipulation.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2022, the patient had a revision right total hip arthroplasty, to address nonreducible posterior dislocation.Depuy components were implanted during this procedure.The indications for surgery included, subluxation episodes and a dislocation.The cup and stem were noted to be ¿perfectly positioned¿ and were retained.The surgeon reported that the patient has deficiency of the posterior soft tissues.During the procedure, the surgeon observed acetabular cup had wear posteriorly, due to being dislocated.While implanting the new liner, the surgeon used a rongeur and ¿just trimmed down the 10-degree portion of the lip just a little bit, to give a little more clearance in extension¿.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Available xray evidence was reviewed.Based in attached xray images it cannot be determinate that cup presented wear.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H10 additional narrative: added: b1 (is product problem) and b5 corrected: h6 (device code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE 100 ACET CUP 54MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13505573
Report Number1818910-2022-02641
Device Sequence Number1388552
Product Code LPH
UDI-Device Identifier10603295008569
UDI-Public10603295008569
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-01-054
Device Catalogue Number121701054
Device Lot Number718139
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2022
Supplement Date Received by Manufacturer03/23/2022
Initial Report FDA Received Date02/10/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/14/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
APEX HOLE ELIM POSITIVE STOP.; BIOLOX DELTA FEMORAL HEAD CERAMIC 36MM +3.; PINNACLE ALTRX 54MM X 36MM +4 NEUTRAL.; PINNACLE SHELL 100 SERIES 54MM.; SROM SLEEVE PROX ZTT 18B LG.; SROM STEM STD 36MM +12L NECK 18.0MM X 13MM X 160MM.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age41 YR
Patient SexMale
Patient Weight86 KG
-
-