| Model Number |
ORT300 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/06/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An imris customer service engineer (cse) was on-site to resolve the issue.A pinhole leak was identified at a sharp bend in the table's hydraulic lines of the lift cylinder.The hydraulic lines to the lift cylinder were replaced and the table was tested to verify proper function.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019, the customer reported that hydraulic fluid leaks occured daily and during surgery from the ort300 table.Approximately 4cm in diameter of puddle fluid leak occurred however; the leak did not affect the function of the table.There was no patient harm or user injury reported due to the event.
|
| |
|