SYNTHES GMBH TFNA FENESTRATED HELICAL BLADE 90MM - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
Catalog Number 04.038.390S |
Device Problem
Migration (4003)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 01/01/2022 |
Event Type
Injury
|
Event Description
|
Device report from depuy synthes reports an event in (b)(6) as follows: it was reported that there was a cut-out of tfna blade.Implant failed.Femoral neck blade has migrated to the center.The issue was identified on (b)(6) 2022.Patient was treated with a hip-tep.The procedure was completed successfully with patient outcome is reported as good.This report is for (1) tfna fenestrated helical blade 90mm - sterile.This is report 1 of 1 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Additional device product code: ktt.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.(b)(4).The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.B3: exact date of event is unknown; event occurred sometime in 2022.Part 04.038.390s, lot 92p9091: manufacturing date: february 23, 2021.Manufacturing site: elmira.A review of the device history record revealed no complaint related anomalies.The device history record shows this lot was processed through the normal manufacturing and inspection operations with no rework or nonconformance's noted.This lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.A review of the device history record revealed no complaint related anomalies.This lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.A photo investigation was completed: visual analysis of the provided photo revealed that the tfna helical blade had migrated out of the bone cortex through the left femoral head.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9; h4 h3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that tfna helical blade perf l90 tan was observed that the complaint device showed normal wear consistent with the device use.X-ray images were provided and show the migration/backout.A dimensional inspection was performed for the tfna helical blade perf l90 tan and met specifications.A functional test was performed with the mating devices (nail) and was able to assemble without any difficulties.The complaint condition was not able to be replicated.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the tfna helical blade perf l90 tan would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.G1: corrected physical manufacturer.
|
|