A steris service technician arrived onsite to inspect the table and found it to be operating properly; no repairs were required.While onsite, the technician was informed by user facility personnel that the x-ray top was not properly secured onto the tabletop at the time of the reported event.As user facility personnel began to articulate the table, the x-ray top became unstable.The 5095 surgical table operator manual states (4-12), "4.13 installing/removing x-ray top.3.Position x-ray top section over corresponding tabletop section, aligning posts to mounting holes.4.Press down on x-ray top section until a "click" is heard.5.Ensure x-ray top sections are securely mounted by gently pulling up on section." a steris account manager performed in-service training on the proper use and operation of the 5095 surgical table, specifically properly locking the x-ray top to the table.No additional issues have been reported.
|