| Model Number |
100/189/070 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Codes |
Airway Obstruction (1699); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was unable to be evaluated as it was not returned.If the device is received in the future an evaluation will be performed.The reported event was unable to be confirmed or replicated.A cause of the reported event could not be determined at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient who was intubated had to be extubated and re-intubated with a non saccet et tube.The et tube had to be replaced due to cuff failure.There was no patient injury reported.
|
| |
|
Event or Problem Description
|
|
Additional info received (b)(6)22: three devices in total were used on the same patient on the same day.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: no lot number was provided; therefore, device history record review could not be performed.A product sample was received for evaluation.Visual inspection found no anomalies in the cuff, but damage was detected in the inflation line.Due the inflation line is damaged, it's not possible perform the leak functional test to confirm the leak in cuff since the inflation line must be connected to leak tester machine to allow air pass through the tube and confirm the leak in cuff.Based on the analysis conducted in the sample provided, the root cause analysis cannot be addressed due to damage in the inflation line.No corrective action required since the reported failure mode was not confirmed.E4 is unknown.Additional information received (updated b5).
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.The returned sample was visually inspected and no anomalies were found in cuff, but damaged was detected in inflation line.Due to the inflation line being damaged, it's not possible perform the leak functional test to confirm the leak in cuff.The inflation line must be connected to leak tester machine to allow air pass through the tube and confirm the leak in cuff.The root cause of the reported issue was not able to be determined.Actions were taken to mitigate the reported issue: no corrective action required since the reported failure mode was not confirmed.No problems or issues were identified during this device history record review.
|
| |
|