| Model Number |
85056 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reportable based on device analysis completed on january 12, 2022.During a procedure to treat tachycardia a blazer ii htd catheter was selected for use.It was reported that there was no electric potential after the catheter entered body.Catheter connection was normal.The catheter was replaced and the procedure was completed successfully with no patient complications.However, device analysis revealed a circular cut in the distal section over the catheter shaft.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to boston scientific for analysis.Visual inspection revealed a kink in the distal section between ring #2 and ring #3.Additionally it was observed a circular cut in the distal section over the catheter shaft was observed.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.Electrical continuity test was performed and the device is out of specification.The thermistor was found open electrically.Ablation testing cannot be carried out because the thermistor is electrically open.It was confirmed this device met specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.It was confirmed this device met specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.
|
| |
|