| Model Number |
67SN035 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/15/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the cuff does not inflate when checked before changing the needle.The device was returned to smiths medical.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to smiths medical.The pilot balloon was manipulated, and the cuff inflated.After the whole cuff inflated, it was deflated and inflated 4 times, all the times the cuff inflated completely and symmetrically, based on analysis the complaint is not confirmed.Dhr: a device history review was also conducted on the component lot numbers and no discrepancies were found.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to smiths medical.The pilot balloon was manipulated, and the cuff inflated.After the whole cuff inflated, it was deflated and inflated 4 times, all the times the cuff inflated completely and symmetrically, based on analysis the complaint is not confirmed.Dhr: a device history review was also conducted on the component lot numbers and no discrepancies were found.
|
| |
|
Event or Problem Description
|
|
It was reported that the cuff does not inflate when checked before changing the needle.The device was returned to smiths medical.
|
| |
|