|
SMITH & NEPHEW, INC. GII OVAL PATELLA 5 PT. SZ GDE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Model Number |
71440842 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the gii oval patella sizing guide is broken.As this was noticed in a non-surgical environment, there was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device was not returned for evaluation but the pictures were reviewed, and the fracture was confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported incident could be corroborated as it may have occurred due to its extensive use.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|