• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS SCARBOROUGH, INC. BINAXNOW COVID-19 AG CARD HOME TEST; LATERAL FLOW IMMUNOASSAY IVD OF COVID-19

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS SCARBOROUGH, INC. BINAXNOW COVID-19 AG CARD HOME TEST; LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 Back to Search Results
Catalog Number 195-100
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
The consumer reported a false negative result with the binaxnow¿ covid-19 antigen self test performed on (b)(6) 2021.On (b)(6) 2021, pcr confirmation testings generated positive result.Consumer confirmed she was symptomatic.No additional patient information, including treatment and outcome, was provided.
 
Additional Manufacturer Narrative
The investigation is still in progress.A supplemental report will be provided after completion.
 
Additional Manufacturer Narrative
Testing was performed at abbott diagnostics scarborough, inc.On retained kit lot 139811h with internal positive quality control samples and negative quality control swabs.All test results were valid and performed as expected.Additionally, the manufacturing records and quality control release testing was reviewed for test kit part number 195-200 / lot 139811h and test base part number 195-430h / lot 135690.The lot met the required release specifications.A review of the complaints reported as false negative patient results (confirmed and unconfirmed, conflicting results) related to kit lot 139811h showed that the complaint rate is(b)(4).In conclusion, abbott diagnostics scarborough was unable to determine the exact root cause of the reported issue; however, it could possibly be related to issues including the home test user performance or interpretation of the result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BINAXNOW COVID-19 AG CARD HOME TEST
Common Device Name
LATERAL FLOW IMMUNOASSAY IVD OF COVID-19
Manufacturer (Section D)
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
10 southgate road
scarborough ME 04074
Manufacturer Contact
rachel blackwell
10 southgate road
scarborough, ME 04074
6613888803
MDR Report Key13509514
Report Number1221359-2022-00831
Device Sequence Number12093397
Product Code QKP
UDI-Device Identifier10811877011337
UDI-Public01108118770113371721091910139811H
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EUA203107
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/19/2021
Device Catalogue Number195-100
Device Lot Number139811H
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2022
Supplement Date Received by Manufacturer05/27/2022
Initial Report FDA Received Date02/10/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-