COOK INC FLEXOR ANSEL GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G29981
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 01/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, during an unknown leg procedure, a flexor ansel guiding sheath leaked "around" the device.A hematoma developed at the puncture site.The patient did not require any additional procedures due to this occurrence.Additional information has been requested.
 
Event or Problem Description
Information was available and inadvertently omitted from the initial report.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Corrected information: b5.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Event summary: as reported, during an unknown leg procedure, a flexor ansel guiding sheath leaked "around" the device.A hematoma developed at the puncture site.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Investigation - evaluation: reviews of the complaint history, device history record, instructions for use, manufacturing instructions, and quality control procedures and a visual inspection of the device were conducted during the investigation.One used flexor ansel guiding sheath (kcfw-5.0-18/38-45-rb-anl1-hc) was received for investigation.No leakage was noted during a leak test.A document-based investigation evaluation was performed.No related non-conformances were recorded, and there have been no other reported complaints for this lot number.The device history record review provides objective evidence that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.A review of relevant manufacturing documents was conducted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The device is provided with instructions for use which include hematoma as a potential adverse event.The ifu also states, ¿upon removal from package, inspect the product to ensure no damage has occurred.¿ based on the available information, cook has concluded that a component failure unrelated to manufacturing contributed to this incident.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR ANSEL GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key13509589
Report Number1820334-2022-00199
Device Sequence Number4064156
Product Code DYB
UDI-Device Identifier00827002299818
UDI-Public(01)00827002299818(17)241018(10)14285611
Combination Product (Y/N)N
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/18/2024
Device Model NumberG29981
Device Catalogue NumberKCFW-5.0-18/38-45-RB-ANL1-HC
Device Lot Number14285611
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/27/2022
Supplement Date Received by Manufacturer02/02/2022
05/09/2022
Initial Report FDA Received Date02/10/2022
Supplement Report FDA Received Date03/04/2022
05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-