| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that non-conforming goods were discovered in the distributor's warehouse in (b)(6) 2021.Subsequently, the issue has been identified as foreign material in the package.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint (b)(4).Concomitant medical products: item sp-st5-6704, lot # 221500, sterile trac system sterile trac 5 xd thinflap 1.5 x 4.0mm screw.Report source: foreign: (b)(6).A visual inspection was conducted on the returned product per inspection criteria.The packaging was then opened to further investigate the total pieces of debris and a single hair could be seen inside the sterile pack.The complaint is confirmed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Root cause is attributed to a manufacturing issue.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2022-00051.
|
| |
|