Product complaint number: (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, and the following was obtained: could you please confirm if the 4 sutures broke? i have enclosed attached photographs of the broken sutures.When was the each suture broke (in the package, during removal from package, during handling or during use on the patient)? please specify the first photo was taken (b)(6) , 2021.It broke upon removal of the package.The second and third photos in this email are the other three broken sutures that occurred on (b)(6), 2021.They were broken while the doctor was using them.You can see from the last photo that all three broken sutures came from the same box/lot.Was there any adverse consequence associated with the patient? no to both individual instances.Have any of these events where the suture snapped prematurely been previously reported to ethicon? if yes, provide the respective reference number(s).No i have not previously reported tis problem before, however, it does some to happen quite frequently with the 6-0 prolenes.If not, please create a product complaint for each event and provide the respective reference number(s).N/a.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent related to: 2210968-2022-00986, 2210968-2022-00987, and 2210968-2022-00988.
|