Model Number 71352366
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Unspecified Infection (1930)
Date of Event 01/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after a thr surgery had been performed on (b)(6) 2021, the patient experienced a large ongoing wound infection that extended into the patient's hip joint.Previous washouts had been performed, but the infection was unable to be cleared.Therefore, revision surgery was carried out on (b)(6) 2022 in order to solve this adverse event.The redapt modular shell 66mm was explanted.The current health status of the patient is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The pictures provided were reviewed and could not confirm the stated failure mode.The clinical/medical investigation concluded that, the clinical root cause of the wound infection, washouts and subsequent revision were reported to have been related to a ¿large ongoing wound infection.¿ it cannot be concluded the reported events were associated with a malperformance of the implant or implant failure.The patient impact beyond the infection and revision cannot be determined.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for total hip systems revealed that infection has been identified as a potential complication associated with total hip arthroplasty.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the sterilization records revealed the batch was sterilized within normal parameters.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDAPT MODULAR SHELL 66MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13510742
Report Number1020279-2022-00653
Device Sequence Number14549303
Product Code LPH
UDI-Device Identifier00885556657898
UDI-Public00885556657898
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K182109
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71352366
Device Catalogue Number71352366
Device Lot Number19HM10188
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2022
Supplement Date Received by Manufacturer03/09/2022
Initial Report FDA Received Date02/10/2022
Supplement Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/13/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
71354034/REDAPT MOD SLV SM SZ 18-19 STIKTITE W/HA
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-