MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 AXIUM HELIX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 AXIUM HELIX; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number QC-2-8-HELIX
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
(b)(6) 2022 (b)(4) (rep, for): medtronic received a report that the spring coil could not be detached after 3 times of normal detachment.It was unknown if the physician attempted to detach via another instant detacher or via a hypotube.The devices were prepared as indicated in the ifu.There were no patient symptoms or complications associated with the event.The patient was undergoing surgery for treatment of an unruptured aneurysm with a max diameter of 4mm and a 3mm neck diameter.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product analysis: as found condition: the axium coil was returned for analysis within a shipping box; within a plastic bio-pouch; and within a dispenser coil.Visual inspection/damage location details: the axium coil was returned without introducer sheath.The axium pushwire was found to be broken from proximal end.The actuator interface, positive load indicator and break indicator were found to be broken/missing and retained by release wire.The missing parts were not returned, and reason was not provided.The axium pushwire was appeared to be bent from ~22.5cm and ~29.8cm from proximal end.A break at this location is indicative that a manual detachment was attempted.As the actuator interface and coupler tube were not returned, any mechanical detachment attempts using an instant detacher cannot be confirmed.The axium implant coil appears to be still attached to the pusher, however, the implant coil was found damaged.Under the microscope, the coin was found to be located against the lumen stop, the shield coil was found to be present.No damage was found with the shield coil.No other anomalies were observed.¿ testing/analysis: due to its damaged condition the axium coil could not be used for detachment testing with an in-house id.¿ conclusion: based on the device analysis and reported information, the customer¿s report of ¿non-detachment¿ was confirmed as the axium implant coil was returned still attached to the pusher.In this event, the damages found with the pusher (bends) likely contributed to the event causing the implant coil to not detach.Damage to the implant coil and pusher can occur if the user advances the coil against resistance.However, the cause for the damage could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported that there were no issues prior to the coil non-detachment, and no pushwire damage was observed.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM HELIX
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key13510858
Report Number2029214-2022-00202
Device Sequence Number5299136
Product Code KRD
UDI-Device Identifier00847536029491
UDI-Public00847536029491
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/22/2024
Device Model NumberQC-2-8-HELIX
Device Catalogue NumberQC-2-8-HELIX
Device Lot NumberB254397
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer02/14/2022
04/10/2022
Initial Report FDA Received Date02/10/2022
Supplement Report FDA Received Date03/10/2022
04/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
-
-