|
It was reported that during a knee procedure, the 4.5 solid drill broke when first used in the knee, little or no force was applied at the time; it broke clean in two pieces and the broken pieces were removed from the patient.The procedure was successfully completed with non-significant delay using a back-up device.No further complications were reported.
|
|
H3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A visual inspection of the returned device found that it is not in its original packaging.The markings on the device confirm the product identification information.The device is worn from use, and it is fractured near the distal end.A review of the customer provided image found that the markings on the device confirm the product identification information.The device is broken near the distal end.According to the report, the procedure was successfully completed with non-significant delay using a back-up device.Since no other complications were reported, no further clinical/medical assessment is warranted at this time.Should any additional relevant medical information be provided, this case would be re-assessed.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, attempted correction of a damaged device, or an impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|