• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NAVIGATOR HD; ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NAVIGATOR HD; ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number M0062502230
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a navigator hd access sheath device was used in the ureter during a ureteral calculus procedure performed on (b)(6) 2022.During unpacking, it was found that the sheath catheter was fractured.The procedure was completed with another navigator hd access sheath device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a navigator hd access sheath device was used in the ureter during a ureteral calculus procedure performed on (b)(6) 2022.During unpacking, it was found that the sheath catheter was fractured.The procedure was completed with another navigator hd access sheath device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h2 (additional information): block b5 has been updated based on the additional information received on september 19, 2022.Block h6: medical device code a0414 captures the reportable event of sheath fractured.Block h10: investigation result the returned navigator hd device was analyzed, and a visual evaluation noted that the sheath was found torn in the middle.Microscopic examination was performed and found that the sheath that was torn had bending marks at the proximal end.No other problems with the device were noted.The reported event of sheath catheter fracture was confirmed according to the evidence, it is possible that operational factors, such as handling/manipulation during preparation could have caused these failures (sheath torn and bending marks) and consequently affect the performance of the device.Therefore, the most probable root cause is adverse event related to procedure.
 
Event or Problem Description
It was reported to boston scientific corporation that a navigator hd access sheath device was used during a ureteroscopy procedure performed on january 10, 2022.During unpacking, it was found that the sheath catheter was fractured.The procedure was completed with another navigator hd access sheath device.There were no patient complications reported as a result of this event.Additional information received on september 19, 2022: the type of procedure performed on january 10, 2022 was ureteroscopy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIGATOR HD
Common Device Name
ENDOSCOPIC ACCESS OVERTUBE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key13513644
Report Number3005099803-2022-00709
Device Sequence Number4158485
Product Code FED
UDI-Device Identifier08714729837985
UDI-Public08714729837985
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K140323
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2023
Device Model NumberM0062502230
Device Catalogue Number250-223
Device Lot Number0027417734
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/13/2022
Supplement Date Received by Manufacturer08/31/2022
Initial Report FDA Received Date02/11/2022
Supplement Report FDA Received Date09/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexMale
Patient Weight55 KG
-
-