MAUDE Adverse Event Report: INVIVO, A DIVISION OF PHILIPS MEDICAL SYSTEMS XPER FLEX CARDIO PHYSIOMONITORING SYSTEM; 453564634201

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVIVO, A DIVISION OF PHILIPS MEDICAL SYSTEMS XPER FLEX CARDIO PHYSIOMONITORING SYSTEM; 453564634201 Back to Search Results
Model Number XPER FLEX CARDIO FC2010 REV C
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Philips technical support (ts) was able to remotely assist the customer and confirm the reported problem via the telephone.According to case notes, the customer replaced the device in question with a spare unit and the issue was resolved.As a result, a replacement device was ordered for the customer.A search in salesforce found no further related calls.
 
Event or Problem Description
The customer reported that the flex cardio 2010 device is having issues when displaying vitals during cases.According to the customer, vitals are skipping or repeating.The device was in use on a patient.There was no report of patient or user harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPER FLEX CARDIO PHYSIOMONITORING SYSTEM
Common Device Name
453564634201
Manufacturer (Section D)
INVIVO, A DIVISION OF PHILIPS MEDICAL SYSTEMS
12151 research parkway
orlando FL 32826
Manufacturer (Section G)
INVIVO, A DIVISION OF PHILIPS MEDICAL SYSTEMS
12151 research parkway
orlando FL 32826
Manufacturer Contact
tara mackinnon
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key13513964
Report Number1051786-2022-00013
Device Sequence Number4159056
Product Code MWI
UDI-Device Identifier00884838084902
UDI-Public00884838084902
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K101571
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberXPER FLEX CARDIO FC2010 REV C
Device Catalogue Number453564634201
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Initial Report FDA Received Date02/11/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/24/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1