(b)(4).Batch # unknown.Product was not returned.Based on the information available, it has been determined that no corrective and preventive action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.The lot/batch was not provided; therefore, a manufacturing record evaluation could not be performed.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|
|
It was reported that the surgeon tried to use the tx30g device to perform the distal transection at the sigmoid colon to remove a tumor during a low anterior resection case.Surgeon is a senior surgeon with many years of experience of using the tx30g device.The tissue was placed comfortably within the jaws of the device, and the device was then closed and fired.However, the two rows of staples were not formed and there was no evidence of the staples even being present in the staple housing.Surgeon then proceeded to use sutures instead to complete the distal transection.The case was delayed by 90-minutes.The case was successfully completed with no fragments or patient consequences.
|
|
(b)(4)date sent: 03/17/2022 d4: batch # u5dz10 device analysis: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the tx30g device arrived with no apparent damaged and with a reload loaded on the device.The reload was received fully fired.The device was tested for functionality with a test reload xr30g, l5458m and it fired and formed all the staples as intended.The staple line was complete, and the staples meet the staple form release criteria as part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: squeeze the closing trigger until a click is heard.The instrument is in an intermediate position, the pin is fully seated in the anvil capturing the tissue, and the jaws are partially open.Reposition tissue within the instrument if desired.Squeeze the closing trigger and the handle fully together until a second click is heard.The closing trigger is now latched to the handle and the jaws are clamped onto the tissue, which is ready to be stapled.The firing trigger will simultaneously move down to the ready-to-fire position.Before firing, check to ensure the retaining pin is seated in the anvil.If the pin is not properly positioned, staples may not form properly, which may result in leakage or disruption of the staple line.Fire the instrument by pulling the firing trigger back completely against the closing trigger until a click is heard, signifying the staples are fully formed.The event described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|