BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941816400
Medical Device Problem Codes Obstruction of Flow (2423); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent moved on balloon occurred.A 4.00 x 16mm synergy xd drug-eluting stent was selected for use.However, it was noted that the stent was not placed on the balloon correctly.No patient complications were reported.
 
Additional Manufacturer Narrative
The synergy xd mr us 4.00 x 16 mm stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal marker bands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues.A visual and tactile examination of the hub/manifold found a hard, plastic-like material stuck inside the hub occluding the connection.An attempt was made to remove the plastic- like material however it was stuck in the hub cavity and could not be removed.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that stent moved on balloon occurred.A 4.00 x 16mm synergy xd drug-eluting stent was selected for use.However, it was noted that the stent was not placed on the balloon correctly straight from package.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13514686
Report Number2134265-2022-01032
Device Sequence Number14801782
Product Code NIQ
UDI-Device Identifier08714729981121
UDI-Public08714729981121
Combination Product (Y/N)Y
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/25/2023
Device Model NumberH7493941816400
Device Lot Number0027096985
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/24/2022
Supplement Date Received by Manufacturer03/03/2022
Initial Report FDA Received Date02/11/2022
Supplement Report FDA Received Date03/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-