ZIMMER BIOMET, INC. RINGLOC BI-POLAR 28X46MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported the bipolar liner would not seat in the bipolar shell.No additional information.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, d4, d9, g3, h2, h3, h4, h6 complaint sample was evaluated and the reported event was confirmed.A ringloc bi-polar 28x46mm was returned and evaluated against the complaint.Visual inspection found the ring to be bent within the locking groove of the shell.The ring is oriented in the appropriate direction with the chamfer facing outward.The ring does not rotate freely within the locking groove.The ring was removed from the shell during the evaluation.No damage was observed to the locking groove of the shell.The liner was found to have scratches and indentations in multiple locations around the outer radius.No damage was observed to the locking groove of the liner.The head rotates freely within the liner.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RINGLOC BI-POLAR 28X46MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13515078
Report Number0001825034-2022-00329
Device Sequence Number19873429
Product Code JDI
UDI-Device Identifier00880304009080
UDI-Public(01)00880304009080(17)260715(10)091070
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K051569
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/15/2026
Device Model NumberN/A
Device Catalogue Number11-165216
Device Lot Number091070
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer04/01/2022
Initial Report FDA Received Date02/11/2022
Supplement Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexFemale
Patient Weight41 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-