MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II CR FEM OX NP RT SZ 8; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II CR FEM OX NP RT SZ 8; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74021158
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 07/01/2015
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after a tka surgery was performed the patient experienced on the right knee pain and stiffness, the issue was resolve by manipulation under anesthesia.The outcome of the patient was resolved with residual effects.
 
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Event or Problem Description
It was reported that after a tka surgery was performed the patient experienced on the right knee pain and stiffness, the issue was resolve by manipulation under anesthesia.The outcome of the patient was resolved with residual effects.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, per complaint details an mua (manipulation under anesthesia) was performed due to right knee pain and stiffness an unknown duration of time post tka.The provided crfs were reviewed; however, do not provide insight into the reported right knee pain and stiffness.Per the crfs, there were no technical difficulties during the procedure performed on (b)(6) 2015.Arthrofibrosis is a known possible post-surgical occurrence, and the provided documentation does not support a component mal-performance.Reportedly, the ¿outcome: resolved with residual effects¿.The root cause of the reported events could not be definitively concluded based on the limited information provided.The patient impact beyond that which was reported cannot be determined.No further medical investigation could be rendered at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for knee systems revealed that stiffness has been identified as a possible adverse effects.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, joint tightness, material in use, patient reaction, alignment, fit/size of device used, wear or loss of sterility.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II CR FEM OX NP RT SZ 8
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13515555
Report Number1020279-2022-00658
Device Sequence Number4474287
Product Code JWH
UDI-Device Identifier00885556422816
UDI-Public00885556422816
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K121443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/24/2025
Device Model Number74021158
Device Catalogue Number74021158
Device Lot Number15DM18039
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2022
Supplement Date Received by Manufacturer03/10/2022
Initial Report FDA Received Date02/11/2022
Supplement Report FDA Received Date03/13/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/27/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age72 YR
Patient SexUnknown
Patient Weight84 KG
-
-