ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012451-20
Medical Device Problem Codes Inflation Problem (1310); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual, functional and scanning electron microscopy (sem) inspections/analysis were performed on the returned device.The reported inflation issue was confirmed.A review of the lot history record identified one manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The balloon was ultimately sent to the sem lab for further analysis.The sem report determined the balloon failure may be attributed to mechanical damage to the outer surface.The leak was located at a fold/crease on the distal shoulder of the balloon.Mechanical damage and shredding were documented at and adjacent to the leak on the od surface.The investigation determined a potential product quality issue based on the elhr exception review, the noted pinhole in the balloon proximal to the distal balloon marker and the confirmed inflation issue during return analysis.The issue is being addressed per internal operating procedures.Abbott vascular will continue to trend the performance of these devices.
 
Event or Problem Description
It was reported that the procedure was to treat a non-calcified, moderately tortuous left anterior descending (lad) coronary artery.The 3.5x20 mm nc trek balloon dilatation catheter (bdc) was being used for post-dilatation; however, the balloon only partially inflated at 11 atmospheres and three inflations.Another nc trek was able to be inflated to 20 atmospheres without issue to post dilate the stent.There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.Returned device analysis identified a pinhole in the balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13515917
Report Number2024168-2022-01445
Device Sequence Number4158560
Product Code LOX
UDI-Device Identifier08717648151996
UDI-Public08717648151996
Combination Product (Y/N)N
Initial Reporter CountryEG
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number1012451-20
Device Catalogue Number1012451-20
Device Lot Number91121G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/25/2022
Initial Report FDA Received Date02/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/21/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age47 YR
Patient SexMale
Patient Weight110 KG
-
-