| Model Number |
INS-0352 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The sheath separated at the handle )next to the black handle).When the rt tried to place a fiducial marker, the brush was no longer able to "throw" and therefore could not place the fiducial.Procedure was completed.Note: this ae is in connection with the same procedure as the 3007222345-2022-00004 ae.During the procedure, two of the same model device, with two different lot numbers, had the same issue.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information regarding the event, the device evaluation, and information that was inadvertently not provided in the initial medwatch report.G3 was blank on the initial report and should have been jan.12, 2022.The device history record was reviewed and no issues were identified that could have caused or contributed to the reported issue.The instrument was returned and it was confirmed that the sheath had come detached from the handle.Based on the results of the investigation, there will be a sheath anchoring and adhesion design change to address this issue.Veran will continue to monitor field performance for this device.This supplemental report is being submitted as a corrective action following a retrospective review of complaints in response to an observation from an external inspection.
|
| |
|
Event or Problem Description
|
|
Correction to typing errors in b5 of the initial report: the sheath separated at the handle (next to the black handle).There was reportedly a two-minute delay to switch out the instruments.There was no patient or procedure impact due to the reported issue.
|
| |
|
Search Alerts/Recalls
|