| Model Number |
3T SKYRA |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Laceration(s) (1946)
|
| Date of Event |
04/04/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
There is a known risk that the introduction of ferrous objects into the mr suite can cause damage or injury if they are drawn towards the magnet.All suites are equipped with safety labels informing users and visitors of the hazards intrinsic to the mr environment.The underlying cause of the incident was user error.The service technician failed to verify that the magnet was deployed and that the server on the equipment rack was located in an mr unsafe zone.This mdr is being submitted outside of the required timeframe as part of remedial action initiated by the manufacturer, in response to internally identified issues regarding failed electronic submissions through webtrader.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, a siemens service technician was working on a software upgrade in the equipment room positioned adjacent to the magnet in the diagnostic room.As the technician was attempting to remove and replace a computer server from the equipment rack, the server attracted to the magnet.The technician's left hand was caught between the server and the magnet.The technician was able to extricate his left hand from between the magnet and the equipment, and damage to the technician's middle finger occurred during this process.The technician was treated in the emergency room onsite.The magnet was ramped down to retrieve the server from the side of the magnet.
|
| |
|