• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS SPECTRA CONCEALABLE PENILE PROSTHESIS; PROSTHESIS PENILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS SPECTRA CONCEALABLE PENILE PROSTHESIS; PROSTHESIS PENILE Back to Search Results
Model Number 720074-03
Medical Device Problem Codes Fracture (1260); Mechanical Problem (1384)
Health Effect - Clinical Code Pain (1994)
Date of Event 11/05/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The implant date, lot number, manufacture date and, expiration date were unable to be obtained through good faith efforts.The data was obtained through a review of the surgical history previously provided by the physician.Investigation summary: based on the information available, the cause that contributed to the reported cylinder anomaly cannot be established as the product is not available for analysis.Device history record (dhr): the dhr confirmed that the device met all material, assembly, and performance specifications.Device technical analysis: the device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Labeling review: a review of the spp instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.The patient symptom of pain found to be listed in the ifu.Investigation conclusion: based on the information available, a conclusion code of cause not established was assigned to this investigation.
 
Event or Problem Description
It was reported that a patient implanted with this spectra penile prosthesis experienced pain in the penis base at the right side, which led to a surgical intervention approximately two weeks after symptoms started.During the procedure, the right cylinder was noted to be broken.The entire device was replaced with an inflatable penile prosthesis.Surgery outcome was reported as successful.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS SPECTRA CONCEALABLE PENILE PROSTHESIS
Common Device Name
PROSTHESIS PENILE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
4089353452
MDR Report Key13516782
Report Number2124215-2022-04392
Device Sequence Number12074266
Product Code FAE
UDI-Device Identifier00878953005294
UDI-Public00878953005294
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K090663
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/12/2021
Device Model Number720074-03
Device Catalogue Number720074-03
Device Lot Number0157471002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/17/2021
Initial Report FDA Received Date02/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/12/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
Patient SexMale
-
-