• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 SOLITAIRE FR4; CATHETER, THROMBUS RETRIEVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 SOLITAIRE FR4; CATHETER, THROMBUS RETRIEVER Back to Search Results
Model Number SFR4-4-40-10
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Rupture (2208); Vascular Dissection (3160)
Date of Event 02/03/2022
Type of Reportable Event Death
Event or Problem Description
Medtronic received information regarding a patient death after a solitaire sfr4 stent detached from a pusher wire during retrieval and was lodged in the vessel.Multiple attempts were made to retrieve the stent and the stent was retrieved successfully but a tear in the vessel ensued.The patient became deceased.After using a different 6x40 stent on first pass it was successfully positioned and then switched to 4x40 stent in question and the event happened when retrieving stent into a (b)(4) catheter. ancillary devices: (b)(4).
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a patient death after a solitaire sfr4 stent detached from a pusher wire during retrieval and was lodged in the vessel.Multiple attempts were made to retrieve the stent and the stent was retrieved successfully but a tear in the vessel ensued.The patient became deceased.After using a different 6x40 stent on first pass it was successfully positioned and then switched to 4x40 stent in question and the event happened when retrieving stent into a react68 catheter. ancillary devices: react68.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
(b)(6) 2022 e1,1 (rep, hcp, medwatch): additional information received reported the lesion characteristics were internal carotid artery (ica) and m1 complete occlusion.Vessel tortuosity was moderate.The patient pasted away immediate post care.The death was related to a medtronic device.The cause of death was a vessel bleed.The procedure was stopped and all devices removed.The patient was hospitalized prior to passing of heavy drug use and cardiac issues.The device was being used to restore blood flow by removing thrombus for the treatment of acute ischemic stroke in a patient with a large vessel occlusion who first received intravenous tissue plasminogen activator (iv t-pa).The solitaire fractured form the delivery wire while the physician was attempting to retrieve the device of the fourth attempt at thrombectomy.Device failed e.G.Broke.The physician was able to retrieve the fractured device piece with a 7mm snare.Immediate angiography showed a catastrophic rupture of the let middle cerebral artery.Open intervention could not be done because the patient had received iv t-pa.The patient was taken to cardiovascular intensive care unit and intubated on cardene drip.
 
Additional Manufacturer Narrative
Event related to medwatch (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLITAIRE FR4
Common Device Name
CATHETER, THROMBUS RETRIEVER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
glen belmer
9775 toledo way
irvine, CA 92618
6122713209
MDR Report Key13517691
Report Number2029214-2022-00206
Device Sequence Number5396631
Product Code NRY
UDI-Device Identifier00763000425678
UDI-Public00763000425678
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K183022
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/18/2024
Device Model NumberSFR4-4-40-10
Device Catalogue NumberSFR4-4-40-10
Device Lot NumberB180491
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer02/16/2022
Initial Report FDA Received Date02/11/2022
Supplement Report FDA Received Date03/08/2022
Date Device Manufactured03/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age36 YR
Patient SexMale
-
-