Model Number PL738SU
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to aesculap ag that a caiman disp.Instr.Non articul.D5/240mm (part # pl738su) was used during a procedure performed on (b)(6) 2022.According to the complainant, during surgery for excision of a mediastinal tumor, a piece at the end of the caiman forceps broke off into the patient.The fragment was retrieved from the patient.The complaint device was returned to the manufacturer for evaluation.An additional medical intervention was necessary.Although requested, additional information has not been made available.The adverse event is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
Additional information - a2, a3.Investigation results: visual investigation: at the upper jaw of the tip a piece of the peek insulation is chipped off.The chipped off piece was enclosed.The test and analysis was done using a digital camera.We made a visual inspection of the instrument.Here we found, that a part of the peek insulation of the upper jaw is chipped off.In the next step we tested the clamping and cutting function of the instrument.The clamping and cutting function worked well.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 2(5) x probability of occurrence 1(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a capa is not necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAIMAN DISP.INSTR.NON ARTICUL.D5/240MM
Common Device Name
SEAL & CUT INSTRUMENTS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key13517857
Report Number9610612-2022-00020
Device Sequence Number12046569
Product Code GEI
UDI-Device Identifier04046964693939
UDI-Public4046964693939
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K151696
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/21/2023
Device Model NumberPL738SU
Device Catalogue NumberPL738SU
Device Lot Number52706153
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer02/15/2022
Initial Report FDA Received Date02/11/2022
Supplement Report FDA Received Date03/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age14 YR
Patient SexFemale
-
-