| Catalog Number |
72290042 |
| Medical Device Problem Code |
Unintended Power Up (1162)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a thr procedure, the werewolf fastseal wand kept cutting off when the button was being pressed but it also continually run if it was held upside down without being pressed.The procedure was successfully completed with no delay using a back-up device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference case-(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a thr procedure, the werewolf fastseal wand kept cutting off when the button was being pressed but it also continually run if it was held upside down without being pressed.The procedure was successfully completed with no delay using a back-up device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|