• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL Back to Search Results
Model Number BXA082902A
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.
 
Event or Problem Description
On (b)(6) 2022, the patient underwent treatment of a stenotic lesion in the subclavian artery using a gore® viabahn® vbx balloon expandable endoprosthesis.The sheath access site was the brachial artery (side unknown).The device met resistance while advancing due to stenosis.The physician attempted to withdrawn the device through the sheath.It was noted that the device separated from the catheter during the attempt to withdraw through the sheath.A cut down was performed to removed the device.Another gore device was used to complete the procedure.
 
Additional Manufacturer Narrative
Final report h10/11: section h6 updated to reflect completion of investigation: engineering evaluation: neither clinical images enabling direct assessment of product performance nor the product itself were returned for evaluation.Gore® viabahn® vbx balloon expandable endoprosthesis instructions for use (ifu), for the appropriate region and time-period, was reviewed with respect to the complaint detail and there are applicable statements.The ifu provides instruction that if excessive resistance is felt, the delivery catheter and sheath be removed together as a unit: ¿using fluoroscopic guidance, advance the delivery catheter over the guidewire via the angiographic sheath.Advance cautiously, especially if resistance is felt.If excessive resistance is felt, remove the delivery catheter and sheath together as a unit.¿ the ifu also warns of the potential for stent dislocation when withdrawing the undeployed vbx device back into the sheath: ¿do not withdraw the gore® viabahn® vbx balloon expandable endoprosthesis back into the introducer sheath once the endoprosthesis is fully introduced.Withdrawing the gore® viabahn® vbx balloon expandable endoprosthesis back into the sheath can cause dislocation and/or damage to the endoprosthesis, premature deployment, deployment failure, and/or catheter separation.If removal prior to deployment is necessary, withdraw the gore® viabahn® vbx balloon expandable endoprosthesis to a position close to but not into the introducer sheath.Both the gore® viabahn® vbx balloon expandable endoprosthesis and introducer sheath can then be removed in tandem.After removal, do not reuse the gore® viabahn® vbx balloon expandable endoprosthesis or introducer sheath.¿.
 
Additional Manufacturer Narrative
G3 / g4 updated pma/510(k)number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Common Device Name
ILIAC COVERED STENT, ARTERIAL
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
marcos ayala
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key13518215
Report Number2017233-2022-02737
Device Sequence Number20161568
Product Code PRL
UDI-Device Identifier00733132637553
UDI-Public00733132637553
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P160021
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/19/2024
Device Model NumberBXA082902A
Device Catalogue NumberBXA082902A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/17/2022
Supplement Date Received by Manufacturer05/23/2022
05/23/2022
Initial Report FDA Received Date02/11/2022
Supplement Report FDA Received Date05/23/2022
05/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-