MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 24674
Medical Device Problem Codes Difficult to Remove (1528); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that balloon rupture and removal difficulty occurred.The 50% stenosed target lesion was located in the non-tortuous and non-calcified pulmonary vein.A 10.0 x 20, 75cm mustang balloon catheter was advanced for dilatation.Inflation was performed three times at 14 atmospheres for 7 seconds.However, during the third inflation, the balloon ruptured and a pinhole was noted in the middle of the balloon.The physician tried to remove the device normally but could not pull it into the sheath and was stuck.The chest was surgically opened and the shaft of the device was cut to retrieve it.The procedure was not completed as same device was unavailable.No further complications were reported.
 
Event or Problem Description
It was reported that balloon rupture and removal difficulty occurred.The 50% stenosed target lesion was located in the non-tortuous and non-calcified pulmonary vein.A 10.0 x 20, 75cm mustang balloon catheter was advanced for dilatation.Inflation was performed three times at 14 atmospheres for 7 seconds.However, during the third inflation, the balloon ruptured and a pinhole was noted in the middle of the balloon.The physician tried to remove the device normally but could not pull it into the sheath and was stuck.The chest was surgically opened and the shaft of the device was cut to retrieve it.The procedure was not completed as same device was unavailable.No further complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: mustang 10.0 x 20, 75cm was received for analysis.The device was received in two sections due to a complete shaft break and balloon circumferential tear.The 6rf sheath used by the customer was returned with the device.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Two complete balloon circumferential tears were identified.The first tear was located at the proximal balloon sleeve.The second tear was located approximately 20mm distal of the first.Both sections of balloon material were noted to have been pushed distally towards the tip of the device.The type of damage is consistent with resistance being encountered and excessive force being applied when withdrawing the device through the sheath/guide.No other issues were noted with the balloon material.A visual examination observed damage to the tip of the device.This type of damage is consistent with resistance being encountered and excessive force being applied to the device.A visual and tactile examination found the shaft of the device to have completely detached at the proximal balloon bond.This type of damage is consistent with resistance being encountered excessive tensile force being applied to the device when withdrawing it through the sheath.A visual examination found that both markerbands had moved on the shaft of the device.This most probably occurred due to damage to the shaft.
 
Event or Problem Description
It was reported that balloon rupture and removal difficulty occurred.The 50% stenosed target lesion was located in the non-tortuous and non-calcified pulmonary vein.A 10.0 x 20, 75cm mustang balloon catheter was advanced for dilatation.Inflation was performed three times at 14 atmospheres for 7 seconds.However, during the third inflation, the balloon ruptured and a pinhole was noted in the middle of the balloon.The physician tried to remove the device normally but could not pull it into the sheath and was stuck.The chest was surgically opened and the shaft of the device was cut to retrieve it.The procedure was not completed as same device was unavailable.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13518423
Report Number2134265-2022-00803
Device Sequence Number14764498
Product Code FGE
UDI-Device Identifier08714729793601
UDI-Public08714729793601
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/15/2023
Device Model Number24674
Device Catalogue Number24674
Device Lot Number0026190446
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/20/2022
Supplement Date Received by Manufacturer02/25/2022
Initial Report FDA Received Date02/11/2022
Supplement Report FDA Received Date03/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE: RADIFOCUS; GUIDEWIRE: RADIFOCUS
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-