| Model Number |
GW1420040X |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the device preparation prior to use, the hydrophilic coating was observed to be peeling.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer for analysis; therefore, the alleged product issue cannot be confirmed.If the device or additional information is received, microvention, inc., will issue a supplemental mdr report.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported complaint is non-verifiable.The returned device is not a microvention product; therefore, the investigation is unable to investigate the complaint as described.Review of the manufacturing record and the shipping inspection record of the involved product code/lot# combination confirmed that there was not any indication of anomaly in them.A search of the complaint file found no other similar report with the involved product code/lot# combination from other facilities.
|
| |
|