Pt info: unknown, information not provided.If implanted, give date: unknown, if lens was implanted, information not provided.If explanted, give date: unknown, information not provided.The intraocular lens (iol) is not returning for evaluation as it remains implanted; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
|
Section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: feb 18, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the handpiece was received with cartridge tip damage, consistent with a handpiece that was dropped.A lens was also observed stuck inside of the cartridge as well as overridden by the plunger rod.The handpiece was disassembled and the assembly was inspected, presenting with no assembly issues.The lens was removed and cleaned, presenting with lens damage, consistent with a lens that was stuck folded in a cartridge.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search of complaints related to this production order (po) was performed.The search revealed that no additional complaints were received for this po.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|