AMO MANUFACTURING NETHERLANDS TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Disturbances (2140); Halo (2227)
Date of Event 01/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the preloaded intraocular lens (iol) was explanted from patient's left eye due to halos and glare from night driving and as the patient experienced visual disturbance.There was no medical or surgical intervention required.It was indicated that the daily activities of the patient were not significantly affected, and the patient has fully recovered.The pre-operation visual acuity was 20/20- and that post-implantation with the lens was 20/40-.Another lens of different model (dcb00) and higher diopter (21.5d) was used as the replacement and the visual acuity after this exchange was 20/60+2.No further information was provided.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation? yes.Section d9: returned to manufacturer: yes.Section d9: returned to manufacturer on: feb 8, 2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed that the lens was received cut into three pieces, which is consistent with a lens that was handled during explant.No issues related to or could have contributed to visual issues were observed.Based on the return condition of the lens, the complaint issue could not be confirmed, and no product deficiency could be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13519043
Report Number3012236936-2022-00374
Device Sequence Number14801695
Product Code HQL
UDI-Device Identifier05050474750692
UDI-Public(01)05050474750692(17)240310
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/10/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU210
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/18/2022
Supplement Date Received by Manufacturer03/08/2022
Initial Report FDA Received Date02/11/2022
Supplement Report FDA Received Date03/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight77 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-