Additional information provided in h.3., h.6.And h.10.The customer indicated the use of a qualified cartridge and handpiece.A viscoelastic was indicated which is not qualified for use with the lens/cartridge indicated.An lens exchange for a difference equal to or greater than two diopters, indicates the event is most likely related to a calculation error.Based on the information provided, the event does not appear to be related to the product.Information was provided that the multifocal, company lens, 12.5 diopter (add power +2.2/+3.2) lens was replaced with a competitors, 16.0 diopter, monofocal lens model with reason visual impairment and a clinical reason mechanical complication.The manufacturer internal reference number is: (b)(4).
|