Model Number TFNT00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Impairment (2138)
Type of Reportable Event Serious Injury
Event or Problem Description
A non-health care professional reported following the intraocular lens implantation the patient has experienced visual impairment hence the lens was explanted.The clinical reason given for the explant was mechanical complication.Additional information has been requested and received stating there was no patient harm.The patient was not hospitalized for the event.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and the documentation indicated the product met release criteria.There have been no other complaints reported in the lot number.The root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in h.3., h.6.And h.10.The customer indicated the use of a qualified cartridge and handpiece.A viscoelastic was indicated which is not qualified for use with the lens/cartridge indicated.An lens exchange for a difference equal to or greater than two diopters, indicates the event is most likely related to a calculation error.Based on the information provided, the event does not appear to be related to the product.Information was provided that the multifocal, company lens, 12.5 diopter (add power +2.2/+3.2) lens was replaced with a competitors, 16.0 diopter, monofocal lens model with reason visual impairment and a clinical reason mechanical complication.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ PANOPTIX TRIFOCAL IOL
Common Device Name
LENS, MULTIFOCAL
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13519091
Report Number1119421-2022-00244
Device Sequence Number11880538
Product Code MFK
UDI-Device Identifier00380652392785
UDI-Public00380652392785
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P040020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/12/2026
Device Model NumberTFNT00
Device Catalogue NumberTFNT00.125
Device Lot Number15214662
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer03/16/2022
Initial Report FDA Received Date02/12/2022
Supplement Report FDA Received Date04/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
B&L LI61A0 16.0; HEALON; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; MONARCH III IOL DELIVERY SYSTEM, INJECTOR
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-