| Model Number |
GEMINUS Volar Distal Radius Plate, Standard, 4 Hole, Right |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation concluded that the device failure may have occured due to an installation error and failure to follow post-op procedure.It was noted that not all screws were installed.As per instructions for use, a non-union or implant failure may occur if not all screws are present.Patient was "very active and not very compliant", and admitted to lifting heavy weights at work.The product was not returned for evaluation.Based on information provided and investigation performed, this does not appear to be a device malfunction.
|
| |
|
Event or Problem Description
|
|
All screws installed in the heads of the plate were fractured post-implantation.Implant collapse and fracture displacement was confirmed via florouscopy 3 months post-implantation.
|
| |
|