Model Number 4002C2457
Medical Device Problem Codes Physical Resistance/Sticking (4012); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat carotid-cavernous fistula (ccf) using penumbra coil 400s (pc400s), a neuron 6f 070 delivery catheter (neuron 070), a px slim delivery microcatheter (px slim), and a guidewire.During the procedure, the physician successfully implanted two pc400s using the neuron 070 and px slim.Next, the physician experienced resistance while advancing the third pc400 through the px slim.Subsequently, the pc400 unintentionally detached with part of the pc400 in the px slim and the other part in the fistula.Therefore, the physician used the guidewire to push the pc400 into the fistula.Next, the physician experienced resistance while advancing a new pc400 through the px slim.The physician decided to retract the pc400; however, while retracting the pc400, it unintentionally detached inside of the px slim.Therefore, the px slim was removed with the detached pc400 inside.The procedure was ended at this point.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 1.3005168196-2022-00052.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key13520948
Report Number3005168196-2022-00051
Device Sequence Number20182683
Product Code HCG
UDI-Device Identifier00814548010809
UDI-Public00814548010809
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/14/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4002C2457
Device Catalogue Number4002C2457
Device Lot NumberF98141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Initial Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age33 YR
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-