• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE PLUS TMZF HIP STEM #5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6021-0537
Medical Device Problem Codes Degraded (1153); Material Erosion (1214); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported: "the doctor says femoral head fell off stem.The stem taper was damaged and the neck of the stem is notched." revision of left hip, size 5 accolade tmzf stem and 40 + 8mm cocr head.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned to the manufacturer.
 
Event or Problem Description
As reported: "the doctor says femoral head fell off stem.The stem taper was damaged and the neck of the stem is notched." revision of left hip, size 5 accolade tmzf stem and 40 + 8mm cocr head.
 
Additional Manufacturer Narrative
Reported event: an event regarding disassociation involving an accolade stem that was mated with an lfit v40 cocr head was reported.The event was confirmed.Method & results: product evaluation and results: the reported device was not returned however photographs were provided for review.Visual inspection of the provided photograph indicated that the devices were covered in blood and tissue consistent with explantation.The trunnion of the stem is severely worn.Damage is consistent with the loss of taper lock.Clinician review: a review of the provided medical information by a clinical consultant indicated: "conclusion of assessment: this inquiry reports the failure of a total hip replacement approximately 11 years after implantation due to femoral head disassociation from the femoral stem with trunnion deformity.I can confirm that this event took place since i was able to review an x ray showing the disassociation and deformity of the trunnion.Regarding the possible root cause of this event, i cannot determine it with certainty.Causes of trunnion deformity and femoral head disassociation are multifactorial including surgical technique factors, patient factors and implant factors." product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to disassociation of the head from the stem.The event was confirmed by visual inspection of provided photographs of the devices and by review of the provided medical information by a clinical consultant.The reported accolade stem was mated with an lfit v40 cocr head.The cocr head has been identified to be within scope of nc and capa.No further investigation for this event is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE PLUS TMZF HIP STEM #5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13521916
Report Number0002249697-2022-00232
Device Sequence Number4158756
Product Code LPH
UDI-Device Identifier04546540510709
UDI-Public04546540510709
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2014
Device Model Number6021-0537
Device Catalogue Number6021-0537
Device Lot Number29023004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/26/2022
Supplement Date Received by Manufacturer04/20/2022
Initial Report FDA Received Date02/14/2022
Supplement Report FDA Received Date05/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/06/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age67 YR
Patient SexMale
Patient Weight96 KG
-
-