MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X50 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X50 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1896-5050S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 12/07/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The customer reported a broken gamma nail which requires revision.Update: non union in patient.Upon return of the devices, one of the locking screw was found broken.
 
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Additional Manufacturer Narrative
Upon completion of the investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
The customer reported a broken gamma nail which requires revision.Update: non union in patient.Upon return of the devices, one of the locking screw was found broken.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the device was returned for evaluation and matches the alleged failure mode.Device inspection revealed the following: the received screw was found to be broken from the mid-shaft region.Although lines of rest are not clearly visible, but they are present on the breakage surface very slightly.Along with that, the smooth surface indicates towards a fatigue fracture due to high cyclic overloading.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the above investigation and the available information, the root cause of the breakage is a user related factor.Non-union and progressive overloading eventually led to the breakage of the nail and subsequently a distal locking screw as well, in a fatigue manner.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
The customer reported a broken gamma nail which requires revision.Update: non union in patient.Upon return of the devices, one of the locking screw was found broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X50 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
marilyne chaumont
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key13523229
Report Number0009610622-2022-00048
Device Sequence Number12090182
Product Code HSB
UDI-Device Identifier04546540202512
UDI-Public04546540202512
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K032244
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/31/2026
Device Model Number1896-5050S
Device Catalogue Number18965050S
Device Lot NumberK0663DD
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer03/08/2022
Initial Report FDA Received Date02/14/2022
Supplement Report FDA Received Date04/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-