• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Unexpected Shutdown (4019)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/21/2022
Event Type  malfunction  
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) had an unexpected shut down.There was no harm or injury to the patient and no adverse event was reported.
 
Manufacturer Narrative
Type of investigation not yet determined: a supplemental report will be submitted upon receipt of additional information.
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) had an unexpected shut down.There was no harm or injury to the patient and no adverse event was reported.
 
Manufacturer Narrative
Type of investigation not yet determined: a supplemental report will be submitted upon receipt of additional information.
 
Manufacturer Narrative
Testing of actual/suspected device (10/3233): a getinge field service engineer (fse) was dispatched to evaluate this unit.The fse was unable to duplicate the reported failure.The fse completed preventative maintenance (pm) with full calibration.The unit passed all functional and safety tests per factory specifications.The unit was returned to the customer and cleared for clinical use.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) was turning on and off.There was no harm or injury to the patient and no adverse event was reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: e1 (site country), h6 (type of investigation, investigation findings and investigation conclusions).Corrected fields: h6 (health effect ¿ clinical code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13523480
MDR Text Key285545063
Report Number2249723-2022-00284
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received02/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/11/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
INTRA-AORTIC BALLOON.; N/A.
-
-