Model Number 700009XT
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
 
Event or Problem Description
Unraveled, lost shape caught in branch per customer.
 
Additional Manufacturer Narrative
A review of the dhr and inspection records was not conducted since a lot number was not provided.One opened device was returned for review.Visual inspection found the device to be covered in dried bodily fluids and the sinusoidal wire to be bent out of shape but was not unraveled.The device was functionally tested and the wire did actuate slowly, due to the presence of dried bodily fluids within the sheath.Evaluation of the device did not confirm a manufacturing error.The most probable cause for the damage to the sinusoidal wire was most likely due to an event within the user environment and not related to the manufacturing of the device.Since a manufacturing error could not be confirmed, no corrective action will be taken at this time.
 
Event or Problem Description
Unraveled - lost shape.Caught in branch per customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEANER XT ROTATIONAL THROMBECTOMY SYSTEM 6F X 65CM
Common Device Name
CLEANER XT
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX 75751
Manufacturer Contact
elnaz rahman
1445 flat creek rd
athens, TX 75751
9036759321
MDR Report Key13523715
Report Number0001625425-2022-00951
Device Sequence Number4477313
Product Code QEW
UDI-Device Identifier00886333222308
UDI-Public00886333222308
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K141617
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number700009XT
Device Catalogue Number700009XT
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer02/08/2022
Initial Report FDA Received Date02/14/2022
Supplement Report FDA Received Date03/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-