• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 11.0MM REAMER HEAD FOR RIA 2 STERILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 11.0MM REAMER HEAD FOR RIA 2 STERILE Back to Search Results
Model Number 03.404.018S
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 01/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2022, while performing a ria, two of the 11mm reamer heads broke and fragments of the splines were retained inside of the patient.The tibia canal was very small and the surgeon commented the bone was very hard.Procedure was completed successfully with a (30) minutes of surgical delay.Fragments were generated and no patient consequences.This report is for (1) 11.0mm reamer head for ria 2 sterile.This is report 1 of 2 for complaint (b)(4).
 
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D4.Expiration date.H6 - codes updated to imdrf codes.Device history lot - manufacturing location: supplier ¿ mark two engineering / inspected and packaged by: monument release to warehouse date: 13-sep-2021 part number: 03.404.018s, 11mm reamer head for ria 2-sterile lot number: 344p040 (sterile) lot quantity: 50 production order traveler met all inspection acceptance criteria.Packaging label log (pll) lppf rev a, lmd rev a was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Scn 20537 supplied by sterigenics was reviewed and determined to be conforming.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Component part(s) reviewed: part number: 03.404.M018, ria ii reamer head 11.0mm lot number: 96p9099 lot quantity: 300 inspection instruction met all inspection acceptance criteria.Redlined inspection sheet, incoming final inspection, ns073690 rev a (cn 161765) to rev b met all inspection acceptance criteria.Certificate of compliance received from mark two engineering dated 28-may-2021 was reviewed and determined to be conforming.Certification for heat treat supplied to mark two engineering from vacu-braze inc.Dated 29-apr-2021 was reviewed and determined to be conforming.Heat treat specified at 50-55 hrc.Results certified at 52 hrc.Certificate of analysis supplied to mark two engineering from banner medical dated 11-sep-2020 was reviewed and determined to be conforming.Raw material certification supplied to banner medical from dunkirk specialty steel dated 13-aug-2020 was reviewed and determined to be conforming.Device history review - 01-feb-2022: dhr reviewed by: mschoenfeld this lot met all dimensional, visual, sterility and packaging criteria with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11.0MM REAMER HEAD FOR RIA 2 STERILE
Common Device Name
REAMER
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13523730
Report Number2939274-2022-00467
Device Sequence Number4473517
Product Code HTO
UDI-Device Identifier10886982273963
UDI-Public(01)10886982273963
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.404.018S
Device Catalogue Number03.404.018S
Device Lot Number344P040
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/16/2022
Supplement Date Received by Manufacturer03/15/2022
Initial Report FDA Received Date02/14/2022
Supplement Report FDA Received Date03/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
11.0MM REAMER HEAD FOR RIA 2 STERILE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age23 YR
Patient SexFemale
Patient Weight93 KG
-
-