MAUDE Adverse Event Report: COOK INC ADVANCE 14 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 14 LP LOW PROFILE BALLOON CATHETER; LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number G50326
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, during an angioplasty procedure and using a advance 14 lp low profile balloon catheter, the balloon quickly lost pressure due to a pinhole in the balloon material.An inflation device was used.And the inflation pressure did not make it past 6atm.Inflation failed during the first attempt.A 5fr cook sheath of unknown length was also utilized during the procedure.Patient's anatomy was not noted to be tortuous or calcified.The lesion was located in the posterior tibial artery.The balloon was removed through the sheath.The procedure was successfully completed with another device of the same type.No unintended section of the device remained inside of the patient's body.No additional procedures were required due to the occurrence.No adverse effects were reported due to the occurrence.
 
Additional Manufacturer Narrative
This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, during an angioplasty procedure and using a advance 14 lp low profile balloon catheter, the balloon quickly lost pressure due to a pinhole in the balloon material.An inflation device was used.And the inflation pressure did not make it past 6atm.Inflation failed during the first attempt.A 5fr cook sheath of unknown length was also utilized during the procedure.Patient's anatomy was not noted to be tortuous or calcified.The lesion was located in the posterior tibial artery.The balloon was removed through the sheath.The procedure was successfully completed with another device of the same type.No unintended section of the device remained inside of the patient's body.No additional procedures were required due to the occurrence.No adverse effects were reported due to the occurrence.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Summary of event: as reported, during an angioplasty procedure and using a advance 14 lp low profile balloon catheter, the balloon quickly lost pressure due to a pinhole in the balloon material.An inflation device was used.And the inflation pressure did not make it past 6atm.Inflation failed during the first attempt.A 5fr cook sheath of unknown length was also utilized during the procedure.Patient's anatomy was not noted to be tortuous or calcified.The lesion was located in the posterior tibial artery.The balloon was removed through the sheath.The procedure was successfully completed with another device of the same type.No unintended section of the device remained inside of the patient's body.No additional procedures were required due to the occurrence.No adverse effects were reported due to the occurrence.Investigation evaluation: reviews of the complaint history, device history record, drawings, instructions for use (ifu), quality control procedures, and specifications were conducted during the investigation.A visual inspection of the returned complaint device was also conducted.The complaint device was returned to cook for investigation.A physical examination and leak test of the returned device confirmed that the balloon leaked from near the proximal bond, but not from the bond itself.Cook completed a review of the device history record (dhr).The dhr for the reported lot found no related non-conformances.A related subassembly lot found 2 relevant non-conformances; all non-conforming products were scrapped, and the lot is inspected 100%.A review of complaint history records shows no other related complaints associated with the complaint device lot.Cook determined that no nonconforming product exists in house or in the field.Cook reviewed the device master record (dmr) and product labeling, including the ifu, to confirm adequate manufacturing specifications, operation parameters, and the design verification testing of the device are in place to identify and prevent this failure mode prior to distribution.The instructions for use (ifu) provides the following information to the user related to the reported failure mode: warnings do not exceed rated burst pressure.Rupture of balloon may occur.Rupture of balloon may occur.Adhere to balloon inflation pressure parameters in the compliance card insert.Over-inflation may cause rupture of the balloon, with resultant damage to the vessel wall.Use of a pressure gauge is recommended to monitor inflation pressures.Do not use a power injector for balloon inflation.Rupture may occur.Balloon introduction and inflation if balloon pressure is lost and / or balloon rupture occurs, deflate balloon and remove balloon and sheath as a unit.Balloon deflation and withdrawal if resistance is encountered during withdrawal, apply negative pressure with a larger syringe before proceeding.If resistance continues, remove balloon and sheath as a unit.How supplied upon removal from package, inspect the product to ensure no damage has occurred.Cook concluded that a component failure unrelated to the design or manufacture of the compliant device caused this incident.The appropriate personnel have been notified and cook will continue to monitor for similar events.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 14 LP LOW PROFILE BALLOON CATHETER
Common Device Name
LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key13524589
Report Number1820334-2022-00224
Device Sequence Number5248073
Product Code LIT
UDI-Device Identifier10827002503264
UDI-Public(01)10827002503264(17)231103(10)13532873
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K170193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/03/2023
Device Model NumberG50326
Device Catalogue NumberPTAX4-14-170-2.5-20
Device Lot Number13532873
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2022
Supplement Date Received by Manufacturer04/20/2022
Initial Report FDA Received Date02/14/2022
Supplement Report FDA Received Date05/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/03/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5FR COOK SHEATH.
Patient SexUnknown
-
-