• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-DETEK INCORPORATED ELECTRODES, MFE, PRO-PADZ, BIPHASIC; DEFIB/PACING ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-DETEK INCORPORATED ELECTRODES, MFE, PRO-PADZ, BIPHASIC; DEFIB/PACING ELECTRODE Back to Search Results
Model Number 8900-2303-01
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), two sets of electrode pads needed to be replaced during chest compressions, causing unnecessary delays in delivering therapy.Complainant indicated that the clinician obtained a set of paddles to continue treating the patient.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Additional Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Additional Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Event or Problem Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), two sets of electrode pads needed to be replaced during chest compressions, causing unnecessary delays in delivering therapy.Complainant indicated that the clinician obtained a set of paddles to continue treating the patient.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Additional Manufacturer Narrative
The actual propadz electrodes from the event were not available for evaluation.Instead, electrodes from retained samples of the same lot number 4121c were investigated.Evaluation of the electrodes did not reveal any irregularities that would have contributed to the reported event.Review of the photo provided does show the gel not attached to the pad/conductive plate itself.However, this was after use on the patient.There is a significant amount of creasing visible along the conductive plate, which suggests that cpr was performed whilst pressing on the pads themselves.The product labeling specifically warns the users not to conduct chest compressions through the pads, as damage to the pads can occur.Without receipt of the pads or a log from the device utilized with the pads, a detailed investigation could not be conducted to determine the circumstances that led to the gel separation.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRODES, MFE, PRO-PADZ, BIPHASIC
Common Device Name
DEFIB/PACING ELECTRODE
Manufacturer (Section D)
BIO-DETEK INCORPORATED
525 narragansett park drive
pawtucket RI 02861
Manufacturer Contact
525 narragansett park drive
pawtucket, RI 02861
4017291400
MDR Report Key13525218
Report Number1218058-2022-00006
Device Sequence Number20182594
Product Code MKJ
UDI-Device Identifier10847946016415
UDI-Public10847946016415
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K895032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/08/2022
Device Model Number8900-2303-01
Device Catalogue Number8900230301
Device Lot Number4121C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2022
Supplement Date Received by Manufacturer01/24/2022
Initial Report FDA Received Date02/14/2022
Supplement Report FDA Received Date03/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-