• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEWHEMPRO (US) VH-3000; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEWHEMPRO (US) VH-3000; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VASOVIEWHEMPRO (US) VH-3000
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Embolism/Embolus (4438); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.Device discarded.
 
Event or Problem Description
The hospital wanted to share some insight about a less frequent complication of evh, vasoview hemopro vh-3000 which is co2 embolisms.They have seen these a couple of times, and were able to catch it immediately and fix it without having to completely stop harvesting.They found that if you keep the cvp lower than the gas pressure, you should not run into this issue.There was no patient adverse patient effects.
 
Additional Manufacturer Narrative
Trackwise id # (b)(4).A lot history review was not applicable.The product is not returning.A lot number was not provided as this was part of a post made in doc matter, 6 years ago.The device was not returned to maquet cardiac surgery for investigation, therefore no evaluation could be performed.It is not possible to confirm the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEWHEMPRO (US) VH-3000
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key13527267
Report Number2242352-2022-00155
Device Sequence Number14462914
Product Code GEI
UDI-Device Identifier00607567700345
UDI-Public00607567700345
Combination Product (Y/N)N
Initial Reporter StateSC
PMA/510(K) Number
K052274
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVASOVIEWHEMPRO (US) VH-3000
Device Catalogue NumberVH-3000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2022
Supplement Date Received by Manufacturer06/17/2022
Initial Report FDA Received Date02/14/2022
Supplement Report FDA Received Date06/17/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-