Model Number 60000000
Device Problem Insufficient Cooling (1130)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2022
Event Type  malfunction  
Event Description
It was reported that the nurse had an alert 113 (reduced water temperature control) in the arctic sun device. Nurse stated that the patient was not cooling to target, target temperature was 36 c, patient temperature was 37. 4 c, flow rate was 3. 3 l/min. , water temperature was 8. 1 c, system hours were 670. 7, pump hours were 557. 9, chiller temperature was 4. 2 c, and mixing pump command was 100 percent. Nurse stated that they had another device and would send current device to the biomed. Per follow up information received via phone on 26jan2022, nurse stated that the therapy was completed successfully with the second arctic sun device. And stated that they labeled the first arctic sun device to be picked up by the biomed. Also reported that there was no impact to the patient.
 
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN DEVICE
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13527594
MDR Text Key285699089
Report Number1018233-2022-00602
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number60000000
Device Catalogue Number60000000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 02/14/2022 Patient Sequence Number: 1
-
-