Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Disturbances (2140); Halo (2227)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient was 20/20 j1 but never adjusted to glare and halos after the implantation of the synergy intraocular lenses (iol) in both eyes.There was no refractive error, or residual astigmatism.The lenses were exchanged to non-johnson and johnson monofocal lenses.No complications were observed during the exchange procedures described as being routine surgeries.Patient outcome was reported as having no complications.No further information was provided.It was unknown from which eyes the lenses were explanted.This report represents the right (od) and a separate report will be submitted for the left (os) eye.
 
Additional Manufacturer Narrative
Age, weight and ethnicity ¿ unknown, not provided.Date of event is unknown.The best estimate date would be between the date the lens was implanted, (b)(6) 2021, and the date it was reported on january 25, 2022.If explanted; give date: unknown, not provided.Initial reporter email address: unknown, not provided.Attempts have been made to obtain missing information; however, to date, no response has been received.Device evaluation: since product was not returned, the complaint issue reported could not be verified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13529078
Report Number3012236936-2022-00361
Device Sequence Number17153701
Product Code HQL
UDI-Device Identifier05050474750708
UDI-Public(01)05050474750708(17)240331
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU215
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/25/2022
Initial Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-