| Model Number |
IBAL UKA, FEMORAL PEG DRILL, DIA. 260 |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2022, it was reported by a sales representative via sems that an ar-611-11 femoral peg drill was creating flaky metal.This was discovered during a uka sportsplasty procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed, heavy circumferential abrasion marks were observed on the flute od.A likely cause of the damage is prying/leveraging the device against a guide or another instrument during use.
|
| |
|